Following are the symbols and icons used by the Inogen G4 portable oxygen concentrator to communicate the power and battery status of your concentrator. These icons appear on the control panel screen, and it’s a good idea to familiarize yourself with the symbols, particularly the high-priority alert symbols, so you can immediately take action when you see them.

Click here for how to check the charge level remaining on your Inogen G4 battery when it’s not installed.

Inogen G4 battery level indicator symbols

Symbol(s)What it means
Inogen battery empty symbol iconBattery is empty.
Inogen battery 10% symbol iconFlashing symbol. Battery has less than 10% charge remaining.
Inogen battery half full symbol iconBattery has approximately 40% to 50% charge remaining.
Inogen battery full symbol iconBattery is full.
Inogen battery fully charged symbol iconThe battery is fully charged and is charging as necessary to maintain the charge.
Inogen battery charging level between 60% and 70% symbol iconBattery is charging with charge level between 60% and 70%.
Inogen battery is charging with charge level less than 10%Battery is charging with charge level less than 10%.
Concentrator power supply symbol iconConcentrator is operating from an external power source with no battery present.

Inogen G4 battery informational messages

Symbol(s)What it means
Setting X Battery
HH:MM
Default display when operating on battery power. “X” represents the selected flow setting (e.g., Setting 2). “HH:MM” represents the approximate time remaining on the battery charge (e.g., 1:45)
Setting X Charging xx
(or) Battery Full
Default display when operating on an external power supply and the battery is charging. “xx%” represents the percent battery charge (e.g., 86%).
Setting X Battery xx%Default display when the battery is not charging or when the time remaining is not available from the battery.
Charging xx% (or) Battery FullDisplayed when the concentrator is plugged in and being used to charge a battery (not being used for oxygen production). It is normal to see a fully charged battery read between 95% and 100% when external power is removed. This feature maximizes the useful life of the battery.

Inogen G4 high-priority battery alert messages

Symbol(s)What it means
Battery Empty
Attach Plug
Concentrator has insufficient battery power to produce oxygen. Attach external power supply or exchange battery, then restart unit if necessary by pressing On/Off button.
Battery HOTBattery has exceeded temperature limit while concentrator is running on battery power. Concentrator has stopped producing oxygen. If possible, move concentrator to a cooler location, then turn power off and back on. Ensure air intake and outlet vents have clear access and particle filters are clean. If condition persists, switch to a backup source of oxygen and contact your equipment provider.

Inogen G4 medium-priority battery alert messages

Symbol(s)What it means
Battery HOT
Warning
Battery has exceeded temperature limit while concentrator is running on battery power. If possible, move concentrator to a cooler location or power unit with an external power supply and remove battery. If condition persists, contact
your equipment provider.

Inogen G4 low-priority battery alert messages

Symbol(s)What it means
Battery Low
Attach Plug
Battery power is low, with less than 10 minutes remaining. Attach external power supply or power down and insert a fully charged battery.
Check BatteryBattery error has occurred. Check the connection of your battery and ensure that it is properly attached and latched on concentrator. If battery error recurs with same battery, stop using the battery and switch to a new battery or remove battery and operate concentrator using external power supply.
Remove Battery to CoolBattery has exceeded its charging temperature and charging has stopped. The battery will not charge while this alert is present but will begin to charge when the battery temperature returns to the normal operating range. If battery charging is desired sooner, remove the battery from the concentrator and allow it to cool in an open area for approximately 10-15 minutes. Then, re-insert the battery into the Inogen One® G4. If the problem still persists, contact your equipment provider.