Inogen One® G5® control panel diagram

Inogen One G5 control panel diagram
#DescriptionFunction
1ON / OFF ButtonPress once to turn “ON”; Press and hold for one second to turn “OFF”.
2Flow Setting Control ButtonsUse the – or + flow setting control buttons to select the setting as shown on the display. There are six settings, from 1 to 6.
3Volume Control ButtonPressing this button will change the volume level, from 1 to 4.
4Audible Alert ButtonPressing this button will toggle the Inogen One® G5’s breath detection audible alert on and off.

Breath Detection Alert Mode. The Inogen One® G5 will alert with audible and visual signals for “no breath detected” when this mode is enabled and no breath has been detected for 60 seconds.

At 60 seconds, the device will enter into auto pulse mode and once another breath is detected, the device will exit auto pulse mode and deliver normally on inspiration. The display’s mode indication area will show a bell icon, flashing yellow light and display message when the alert is enabled.

If power is lost, the breath detection audible alert remains set in the user
preferred mode.
5DisplayThis screen displays information regarding flow setting, power status, battery life and errors. Display appearance will vary. Before use, remove the static cling FCC label from the screen.
6Indicator LightsA green light indicates breath detection. A yellow light indicates either a change in operating status or a condition that may need response (alert). A flashing light is higher priority than non-flashing.
7Audible SignalsAn audible signal (beep) indicates either a change in operating status or a condition that may need response (alert). More frequent beeps indicate higher priority conditions. The default volume is set at level 1 and can be adjusted to higher settings but it can not be silenced. If power is lost, the audible signal remains set in the user preferred adjusted setting.
8BacklightA backlight will illuminate the screen for 15 seconds when the on/off button is briefly pressed.